AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE Back to Search Results
Model Number CF-H180AI
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the colonovideoscope had a damaged supportable sleeve with a cut in the metal braid cable.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24067928
Report Number9610595-2026-04507
Device Sequence Number15790337
Product Code FDF
UDI-Device Identifier04953170339585
UDI-Public04953170339585
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCF-H180AI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/23/2025
Initial Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/24/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-