| Model Number |
7700FR38 |
| Medical Device Problem Codes |
Material Separation (1562); Adverse Event Without Identified Device or Use Problem (2993); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Prolapse (2475); Mitral Valve Insufficiency/ Regurgitation (4451); Insufficient Information (4580)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that 4 years and 10 months post implant of this 38mm mitral annuloplasty ring, it was explanted and replaced with a 38mm mitral annuloplasty heart band.The reason for the replacement was not reported.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated data: a.4: weight in lbs: b.1: adv evnt/prod prob: b.2: outcome attributed to adverse event b.5: event description h.6: coding medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that reported the reason for replacement as dehiscence of the annuloplasty band and a p2 leaflet prolapse, causing severe mitral regurgitation.No additional adverse patient effects were reported.
|
| |
|