| Model Number |
A209 |
| Medical Device Problem Codes |
No Audible Alarm (1019); Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this s-icd was thoroughly inspected and analyzed.A high current drain was detected.Detailed analysis confirmed the identified high current drain was a result of a compromised capacitor.This resulted in the observed premature battery depletion.It was determined that the capacitor was compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of emblem devices that have a potential of exhibiting this behavior.This particular device is included in the emblem s-icd accelerated depletion advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) could not be interrogated.No tones were emitted when a magnet was placed over the device.Technical services (ts) discussed troubleshooting options including attempting with a new programmer.This device has been in service approximately nine years and will be replaced.No adverse patient effects were reported.Later this s-icd was explanted for normal battery depletion with successful replacement.
|
| |
|