| Model Number |
L100 |
| Medical Device Problem Codes |
Data Problem (3196); Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker exhibited an alert message indicating that battery replacement indicator had been reached on january 1, 2000 and that remote monitoring was disabled.Analysis of device data was requested.Technical services discussed that this was a false positive explant indicator related to a software update and that all other device functions remain available.Additionally, it was noted that the battery status is at beginning of life and that due to the elective replacement indicator (eri) values not been set, the number zero is causing the date to show as january 1, 2000.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Boston scientific issued a field safety notice update to physicians regarding unintended behaviors observed with a software update designed to detect high battery impedance and prevent initiation of safety mode in accolade devices.The associated investigation determined that the software update led to this device experiencing a false positive response that resulted in disablement of the wandless telemetry.Boston scientific has developed an additional software update to address this unintended behavior.
|
| |
|