| Model Number |
IMPLANT SYSTEM, FIBERTAK BUTTON |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 30-dec-2025, a sales representative reported via phone that three ar-3680 fibertak button implant systems pulled out of the bone while shuttling sutures.Attempts to convert down to the anchor were unsuccessful, as the implants continued to pull out instead of converting.All three implants were subsequently removed intact from the patient.The case was completed using an ar-2260 distal biceps repair implant system.The issue resulted in an approximate 20-minute surgical delay, and it is unknown whether additional anesthesia was administered.The event occurred during a proximal biceps tenodesis procedure on (b)(6) 2025.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|