The device evaluation details.The smart touch unidirectional device was returned to johnson & johnson medtech (j&j medtech) for evaluation on (b)(6) 2025.A visual inspection and screening test of the returned device were performed following j&j medtech procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and no errors were observed.The force feature of the device was evaluated and the force values and the vector were observed within specifications.No force issues were observed.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The reddish material inside the pebax could be related to the force issue reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.In addition, in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.As part of j&j medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number:(b)(4).
|
|
It was reported that a patient underwent a cardiac ablation procedure with a thermocool smarttouch catheter for which biosense webster¿s product analysis lab (pal) identified a hole in the surface of the pebax.The contact of the catheter was not coherent during the procedure.It always detected a contact with a catheter even if it was the only catheter into the atrium.We changed first the cable, a new catheter resolved the issue.We also saw some blood in the rotor of the defective catheter.No patient consequence.The biosense webster, inc.Product analysis lab received the device for evaluation and per the evaluation completion on (b)(6) 2025 observed reddish material and a hole in the surface of the pebax.The event was originally considered non-reportable, however, bwi became aware of a hole in the surface of the pebax on (b)(6) 2025 and have assessed this returned condition as reportable.
|