| Model Number |
000000000000305930 |
| Medical Device Problem Codes |
Difficult or Delayed Activation (2577); Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
12/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|
Event or Problem Description
|
|
The emergency room nurse followed standard procedure when drawing and administering medication.After administering the medication, the nurse attempted to activate the white locking arm to cover the needle, but it can not be pushed.The nurse tried to push harder, but it still wouldn't move, resulting in needle stick.Sample availability: no sample availability details: no sample available.
|
| |
|