MAUDE Adverse Event Report: EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 1.0ML 29GA 1/2IN BLS 400 SG; Syringe, piston

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMBECTA MEDICAL I LLC - HOLDREGE, NE SYRINGE 1.0ML 29GA 1/2IN BLS 400 SG; Syringe, piston Back to Search Results
Model Number 000000000000305930
Medical Device Problem Codes Difficult or Delayed Activation (2577); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 12/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event or Problem Description
The emergency room nurse followed standard procedure when drawing and administering medication.After administering the medication, the nurse attempted to activate the white locking arm to cover the needle, but it can not be pushed.The nurse tried to push harder, but it still wouldn't move, resulting in needle stick.Sample availability: no sample availability details: no sample available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1.0ML 29GA 1/2IN BLS 400 SG
Common Device Name
Syringe, piston
Manufacturer (Section D)
EMBECTA MEDICAL I LLC - HOLDREGE, NE
Manufacturer Contact
anna garbowska-zakrzewski
300 kimball dr.
parsippany, NJ 07054
MDR Report Key24071002
Report Number3024508819-2026-00004
Device Sequence Number18424430
Product Code FMF
UDI-Device Identifier30382903059301
UDI-Public30382903059301
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K992734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000000000000305930
Device Lot Number4218016
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/24/2025
Initial Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
Patient EthnicityNon Hispanic