| Catalog Number |
364390 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd preset¿ eclipse¿, a crack was observed in the hub, resulting in blood leakage, requiring recollection of the sample.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd did not receive any samples or photos for investigation.Therefore, a visual inspection was conducted on 10 retained samples, and no damaged stoppers were observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: damaged.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd preset¿ eclipse¿, a crack was observed in the hub, resulting in blood leakage, requiring recollection of the sample.No adverse impact was reported regarding the patient or user.
|
| |
|