• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC NONE; ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC NONE; ENDOSCOPE Back to Search Results
Model Number 470066-13
Medical Device Problem Code Image Orientation Incorrect (1305)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation was completed to determine the cause of this reported event.An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Based on the field evaluation, this reported event was not confirmed.The fse performed a test drive of the system, but was unable to reproduce the reported issue.The system was tested and verified as ready for use.The complaint was not confirmed based on the field evaluation.The probable root cause cannot be determined based on the information provided and fse's field evaluation.The fse tested the system but was unable to replicate the reported issue.The fse's service visit did not reveal any issues related to the customer reported event.Blank mdr fields: the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Field e4 is blank because it is unknown if the initial reporter submitted a report to the fda.Fields g5 and g7 are not applicable.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the scope image was inverted and flipped, causing left to appear as right and up as down.The caller mentioned that they attempted to swap scopes, but this did not resolve the issue.Ultimately, they rebooted the system, which corrected the orientation problem.The procedure was completing as planned with no reported injury.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) followed up with the initial reporter on and obtained the following additional information: this was not an endoscope issue per the customer.The robot flipped the visual field when rebooted.All endoscopes worked fine after the robot was rebooted.
 
Event or Problem Description
Refer to h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Common Device Name
ENDOSCOPE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key24071756
Report Number2955842-2026-01085
Device Sequence Number9387835
Product Code NAY
UDI-Device Identifier00886874120293
UDI-Public(01)00886874120293
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K232610
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 02/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470066-13
Device Catalogue Number470066
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/23/2025
Supplement Date Received by Manufacturer01/21/2026
Initial Report FDA Received Date01/14/2026
Supplement Report FDA Received Date02/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/28/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES.
Patient SexUnknown
-
-