| Catalog Number |
367859 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® k2 edta (k2e) 5.4 mg blood collection tubes, five (5) tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received any samples or photos for investigation.A total of 100 retained samples were visually inspected, confirming that the hemogard closure assembly was correctly assembled and placed on the tubes.Additionally, functional tests were performed on 10 retained samples.All tubes were found to be within specification limits, with no issues identified.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: draw volume.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® k2 edta (k2e) 5.4 mg blood collection tubes, five (5) tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|