| Catalog Number |
106697 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external blood leak was observed from the access pod of a prismaflex m100 set during continuous renal replacement therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed the reported leak from the set access post-pump pod.This component is manufactured by a vantive external supplier.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the pod defect was determined to be related to supplier manufacturing, and the issue is being further investigated.Nonconformance records have been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|