The manufacturer was contacted in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.A remstar plus device was returned to a third-party service center for evaluation.There was no report of serious or permanent harm or injury.There was no report of medical intervention.During the evaluation of the device, the third-party service center visually inspected the device and found evidence of deteriorated foam.In addition, device display error code e031(err_motor_vbus_high_blower_off).Dust was found as contamination.The device was scrapped.
|