• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 3389S-40
Medical Device Problem Codes Break (1069); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a deep brain stimulation procedure, the lead tip was found to be bent.The lead was never implanted.The surgery was completed on the same day with the replacement of a new lead.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information clarified that the issue occurred out of box.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: analysis of the lead had shown that there was slight damage but this happened during the analysis process.Otherwise there were no issues found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key24073363
Report Number2649622-2026-01231
Device Sequence Number13039902
Product Code MHY
UDI-Device Identifier00763000274405
UDI-Public(01)00763000274405(17)270830(10)VA32FGQ
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3389S-40
Device Catalogue Number3389S-40
Device Lot NumberVA32FGQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer01/14/2026
04/06/2026
Initial Report FDA Received Date01/14/2026
Supplement Report FDA Received Date01/29/2026
04/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11...."
Patient SexUnknown
-
-