| Model Number |
9733235 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3, h6) no products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a sacroiliac and thoracolumbar procedure.It was reported that during a posterior lumbar fusion, the perc pin side pin did not fit into the impaction cap.Two perc pins from the same lot had this issue.The clinical specialist (cs) found a perc pin from a different lot and did not encounter this issue.There was a less than one hour surgical delay.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H2, h3, h6) the two returned perc pins were returned to the manufacturer for evaluation.It was reported that both pins had difficulty inserting into perc pin adapters but were able to insert into perc pin reference frames.The side pin appeared to be oversized.Codes b01, c07, and d02 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|