• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUIDEL CARDIOVASCULAR INC. TRIAGE CARDIAC TNI PANEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUIDEL CARDIOVASCULAR INC. TRIAGE CARDIAC TNI PANEL Back to Search Results
Model Number 97021HS
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Swelling/ Edema (4577)
Date of Event 12/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation conclusion: the customer's complaint was not replicated during in-house testing with retained devices of lot t16083rn.No issues with recovery were observed and the product performed as expected.Manufacturing batch records for the lot were reviewed and found that the lot met release specifications.Unable to determine the root cause of the complaint.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
 
Event or Problem Description
Customer reported discrepant low tni results on triage vs.Vitros.Patient symptoms: sob, swelling of extremities.Treatment: left against medical advice.Patient was not treated based on triage.Diagnosis: swelling, retaining fluid due to severe chf.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAGE CARDIAC TNI PANEL
Common Device Name
TRIAGE CARDIAC TNI PANEL
Manufacturer (Section D)
QUIDEL CARDIOVASCULAR INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
chelsea lindgren
9975 summers ridge road
san diego, CA 92121
MDR Report Key24074996
Report Number3013982035-2026-00001
Device Sequence Number13328815
Product Code MMI
UDI-Device Identifier30014613337935
UDI-Public(01)30014613337935(10)T16083RN(17)260517(11)250903
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K030286
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2026
Device Model Number97021HS
Device Lot NumberT16083RN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2025
Initial Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TRIAGE METERPRO, SN: (B)(6)
Patient SexUnknown
-
-