• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. NEXIVA; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. NEXIVA; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 383539
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
Complaint description: hole in lateral line nexiva needle.Has this product already been used in patients: yes.Has there been a risk to the patient or user: no.
 
Additional Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Additional Manufacturer Narrative
The complaint of damaged extension tubing was confirmed; however, the root cause could not be determined.One 18g nexiva iv catheter was returned for investigation with evidence of use.A portion of the iv catheter was covered in the dressing that was used to secure it.The exposed section of the extension tubing exhibited evidence that it had ballooned or expanded from its original shape.A functional test revealed no leaks or occlusions in any part of the iv catheter.The damage likely occurred during use; however, the root cause could not be verified.This type of damage can occur if the catheter is kinked, obstructed, or pressurized beyond the pressure limit.The instructions for use (ifu) indicate to avoid kinking or obstructing catheter during power injection.In the event of an occlusion, power injector pressure-limiting features may not prevent catheter failure.The ifu also indicates that the 22-18 ga (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kpa) when access ports not suitable for use with power injectors are removed.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Complaints received about this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the data collected for identification of emerging trends.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXIVA
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key24075066
Report Number1710034-2026-00015
Device Sequence Number10266618
Product Code FOZ
UDI-Device Identifier00382903835393
UDI-Public(01)00382903835393(11)250730(17)280731(10)5211924
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K183399
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number383539
Device Lot Number5211924
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Supplement Date Received by Manufacturer02/11/2026
Initial Report FDA Received Date01/14/2026
Supplement Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1