• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 DRIVER; SHOULDER INSTRUMENT - DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 DRIVER; SHOULDER INSTRUMENT - DRIVER Back to Search Results
Catalog Number 620510105
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that the inhance drivers showing plastic wear around reamer attachment and disengaging easily when reaming.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint recon described in this report.H11 additional narrative: added: d4 lot, d9, h4.Corrected: d4 primary udi number.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: it was reported that the inhance drivers showing plastic wear around reamer attachment and disengaging easily when reaming.Replacements was requested to send to the office and in the reference hospital name in the shipping label.The product was returned to depuy synthes for evaluation.Visual analysis found that the distal section of driver have the tabs deformed.Additionally the sample presents small cracks amongst the tabs.The damage reported in this complaint is consistent with the failure mode investigated during a series review of additional complaints involving similar drivers of the same product code (620510105).The series review revealed the cracks of the radel slider handle of the device was loaded in reverse bending beyond the material limit resulting in fatigue fracture initiation and propagation with a small region in the middle of each fracture surface.The potential cause is traced to material overload when the device is spatially oriented in such a way to take on unexpected load during use.A functional test was not performed since the mating device was not returned, however the observed deformed condition can lead with assembling issues.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the driver would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER
Common Device Name
SHOULDER INSTRUMENT - DRIVER
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24077928
Report Number1818910-2026-00900
Device Sequence Number18570187
Product Code HWR
UDI-Device Identifier10603295539421
UDI-Public(01)10603295539421(10)179182
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number620510105
Device Lot Number179182
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2026
Supplement Date Received by Manufacturer01/14/2026
02/09/2026
Initial Report FDA Received Date01/15/2026
Supplement Report FDA Received Date01/20/2026
02/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
-
-