| Model Number |
70914-70 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The useful life of freestyle freedom lite meter is 5 years.As the manufacturing date of this product is 2019, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.The returned meter was investigated with retained test strips and a known-good retained battery.Visual inspection was performed on returned meter.No issues were observed.Meter powered on with button depression and test strip insertion.Blank screen was not observed therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Search Alerts/Recalls
|