MEDTRONIC MEXICO S. DE R.L. DE CV CONCERTO NYLON; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
NV-2-4-HELIX |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that two concerto coils were stuck in the catheter and encountered resistance in the sheath.It was reported that during the procedure, the products became stuck inside the catheter and the coils did not advance.They were replaced with new ones, and the procedure was completed without complications.No patient symptoms or complications were associated with this event.Additional information was received that the coil reportedly became lodged inside the catheter and was removed with difficulty.The cause of the resistance was not determined.A continuous saline flush was administered and maintained as indicated in the ifu.The catheter was not a medtronic device.During preparation, the introducer sheath was held vertically when the implant coils were pulled back inside during hydration.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis as found condition (condition of returned device): two concerto implant coils were returned for analysis within a shipping box; within a sealed biohazard bag; within a resealable plastic bag; within a sealed tyvek biohazard pouch; within their opened concerto implant coil inner pouches and within separate dispenser tracks.Visual inspection/damage location details: the concerto implant coil was returned within introducer sheath.The introducer sheath was found correctly assembled on the pushwire with the wavelock at the proximal end.The concerto coil pushwire was found to be kinked at ~3.2cm, ~70.8cm, and bent within introducer sheath at ~87.9cm from proximal end.The introducer sheath was found to be flattened at distal end.The implant coil was found to be stretched and damaged within introducer sheath.The implant coil was unable to be pushed out further from introducer sheath for additional analysis as it was found to be stuck.No other anomalies were observed.Testing/analysis (including sem reports): the concerto coil tip od (outer diameter) was measured to be 0.0109¿ which is within speci fications.The id (inner diameter) of the introducer sheath was measured to be 0.0185¿ which is within specifications.Conclusion: based on the device analysis and the reported information, the customer¿s report of ¿coil resistance/stuck in sheath¿ was confirmed.However, the root cause could not be determined.There were no reported issues pushing the concerto coil out from within its introducer sheath during preparation.Therefore, it is possible that the implant coil became damaged during return of the implant coil back into the introducer sheath.It is possible insufficient preparation of the concerto coil, damage during preparation or improper hubbing technique (over tightening of the rhv) could have contributed to the event.It is likely the damaged (flattened) introducer sheath for pli-10 contributed to the reported resistance.It is possible the damage to the introducer sheath was cause by over tightening of the rhv.However, the root cause could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|