| Model Number |
8000-004007-01 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, power cycling, and functional stress testing with the returned accessories on a simulator without duplicating the report.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|