• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. LIFELINE; AUTOMATED EXTERNAL DEFIBRILLATORS Back to Search Results
Model Number DDU-100A
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Troubleshooting of the aed with the customer did not identify a cause for the depleted battery pack.The customer was referred to a distributor to purchase a replacement battery and as a follow-up with the customer, the proper maintenance of this device was reviewed.
 
Event or Problem Description
A customer reported their aed is flashing red and prompting a "replace battery pack now" warning.They reported this did not occur during patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT 06405
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key24085106
Report Number3003521780-2026-00028
Device Sequence Number13186997
Product Code MKJ
UDI-Device Identifier10815098020000
UDI-Public(01)10815098020000
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-100A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Initial Report FDA Received Date01/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-