It was reported that the patient presented to a clinic with purulence surrounding a restored implant at tooth site #31.During an examination of this tooth site, bone loss and infection were discovered.The implant was removed with site debridement.After four months of healing, an implant replacement is expected.Symptoms related to event: inflammation, pain and swelling.
|
|
Zimvie complaint number (b)(4).A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|