Getinge became aware of an issue with one of surgical lights - eqt.It was stated, upon the device inspection two out of the six fixation screws holding the device main tube was broken.There was no injury reported, however, we decided to report the issue in abundance of caution as broken screw holding the device configuration may lead to the device detachment from the ceiling and serious injury.
|
|
Getinge became aware of an issue with one of surgical lights - eqt.As it was stated two of fixation screw holding the device main tube was loose.There was no injury reported, however, we decided to report the issue in abundance of caution as loose screws holding the device configuration may lead to the device detachment from the ceiling and serious injury.
|
|
The correction of b5 describe event and problem, e4 occupation and e2 health professional deems required.This is based on the internal evaluation.Previous b5 describe event and problem: getinge became aware of an issue with one of surgical lights - eqt.It was stated, upon the device inspection two out of the six fixation screws holding the device main tube was broken.There was no injury reported, however, we decided to report the issue in abundance of caution as broken screw holding the device configuration may lead to the device detachment from the ceiling and serious injury.Corrected b5 describe event and problem: getinge became aware of an issue with one of surgical lights - eqt.As it was stated two of fixation screw holding the device main tube was loose.There was no injury reported, however, we decided to report the issue in abundance of caution as loose screws holding the device configuration may lead to the device detachment from the ceiling and serious injury.Previous e2 health professional: blank.Corrected e2 health professional: no.Previous e4 occupation: blank.Corrected e4 occupation: biomedical engineer.Getinge became aware of an issue with one of surgical lights - eqt.As it was stated two of fixation screw holding the device main tube was loose.There was no injury reported, however, we decided to report the issue in abundance of caution as loose screws holding the device configuration may lead to the device detachment from the ceiling and serious injury.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.Provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment.Root cause analysis was performed by a subject matter expert at the manufacturer¿s, according to analysis in case of loosening, the probable causes of loosening correspond to an improper initial tightening during the installation or a maintenance not in accordance with the manufacturer's recommendations.The yearly preventive maintenance program mentions: to check the suspension fixing screws.It is also indicated to not retighten the screws during maintenance.If screws appear loosened replace them.To replace the 6 fixing screws every 6 years.In april 2019 getinge transmitted the service bulletin 98a on getinge online reminding to carry out preventive maintenance and wearing parts replacement to ensure the device safety, the service bulletin 98a mentions to replace the suspension fixing screws every 6 years during the preventive maintenance.Moreover, the voluntary fsca-808092, relating the maintenance program on maquet sas or light systems, was communicated in september 2023, in order to focus especially the periodic replacement of the suspension fixing screws.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
|