| Catalog Number |
620510102 |
| Medical Device Problem Codes |
Break (1069); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that pt had left reverse tsa.During glenosphere insertion, the threads on inserter broke in sphere, limiting the ability to advance the locking screw.This added 20 minutes to the case.We had to drill and burr out the broken piece to advance locking screw in a definitive position.A small and lare screwdriver were also damaged during the removal of broken pieces.The broken piece was removed in its entirety.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H11: upon further review, it was determined that the allegation against the star driver is no longer considered a reportable malfunction.Please disregard this mrn.No additional reports will be submitted against it unless further information is obtained that raises the damage caused to a reportable malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|