Model Number WA4KA430
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that during sedation the rigid scope had small crack at distal tip.The issue was found during a diagnostic arthroscopy procedure that was completed with a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: h2, h4, h6, h11.The device was returned to olympus for inspection and the reported failure was not confirmed.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE ¿ULTRA¿, 4 MM, 30°, WITH TROCAR TUBE CONNECTOR
Common Device Name
RIGID SCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24093113
Report Number9610773-2026-00388
Device Sequence Number10214666
Product Code HRX
UDI-Device Identifier04042761083010
UDI-Public04042761083010
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K160249
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA4KA430
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer03/02/2026
Initial Report FDA Received Date01/16/2026
Supplement Report FDA Received Date03/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-