| Model Number |
DDBB1D1 |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/19/2025
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was oversensing, undersensing, and a lead integrity alert (lia) due to high-rate non-sustained episodes and short ventricular intervals on the right ventricular (rv) lead.A charge circuit time out occurred on the implantable cardioverter defibrillator (icd) during a ventricular fibrillation (vf) storm.It was also noted there was noise on the right atrial (ra) lead.The patient passed away three days later.
|
| |
|