• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006L8405920
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the procedure to treat kidney stones, there were five fibers that broke and cracked within minutes of using them.For all fibers, it was noted that this was the first time they had been used.All fibers were broken at the glass tip, however it was reported that no fiber fragments fell inside the patient.The fibers were replaced, and the procedure was completed using another fiber.There were no complications or injury to the patient or users.This report is for the fourth fiber.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The product record review results detailed in the prr table did not determine probable cause.A device history record review (dhr) was performed and confirmed that the device met specification prior to shipping.No manufacturing deviations were noted that could have contributed to the event reported.A labeling review was performed on the device instructions for use (ifu).The ifu cited that, the fiber should not be clamped with forceps or other securing instruments as it could result in fiber damage or breakage.Based on the information available, the conclusion code "unable to exclude device problem" has been assigned as the most probable cause.
 
Event or Problem Description
It was reported that during the procedure to treat kidney stones, there were five fibers that broke and cracked within minutes of using them.For all fibers, it was noted that this was the first time they had been used.All fibers were broken at the glass tip, however it was reported that no fiber fragments fell inside the patient.The fibers were replaced, and the procedure was completed using another fiber.There were no complications or injury to the patient or users.This report is for the fourth fiber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA PULSE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key24095334
Report Number2124215-2026-03206
Device Sequence Number16069483
Product Code GEX
UDI-Device Identifier08714729978589
UDI-Public(01)08714729978589(17)280414(10)36300485
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K210925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM006L8405920
Device Catalogue NumberM006L8405920
Device Lot Number0036300485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2025
Supplement Date Received by Manufacturer01/27/2026
Initial Report FDA Received Date01/16/2026
Supplement Report FDA Received Date02/17/2026
Date Device Manufactured04/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-