| Model Number |
610-2231011-01 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device powered on without display.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was confirmed to be caused by a faulty monitor board.The monitor board was replaced to remedy the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|