• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. GOLD RELOAD FOR ECHELON 60; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. GOLD RELOAD FOR ECHELON 60; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ECR60D
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure, it was observed that the device would not fire.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 1/16/2026.D4: batch #258c74.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecr60d reload was returned unfired with the reload pan partially dislodged from the left proximal side.In addition, 8 double drivers missing, 2 single drivers missing, 1 double driver out of position, making the reload non-functional.No functional test was performed due the condition of the reload.No conclusion could be reached on what may have caused the reload pan to dislodge.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device ecr60d with lot number 314c50; and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 258c74, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOLD RELOAD FOR ECHELON 60
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key24100983
Report Number3005075853-2026-00580
Device Sequence Number10397051
Product Code GDW
UDI-Device Identifier10705036011842
UDI-Public(01)10705036011842(10)314C50(17)280131
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECR60D
Device Lot Number314C50
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2026
Initial Report FDA Received Date01/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-