| Model Number |
419488 |
| Medical Device Problem Codes |
High Capture Threshold (3266); Unstable Capture Threshold (3269)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date is approximate.Month and year are confirmed valid.Continuation of d10: dtma1d1 crt-d implanted: (b)(6) 2023 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that right ventricular (rv) lead trends noted spikes for rv pacing impedance, rv defibrillation impedance, and superior vena cava (svc) defibrillation impedance.An alert also triggered for high out of range svc defibrillation impedance.In addition, left ventricular (lv) lead thresholds were noted as historically high and variable.The leads remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|