• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM MODULAR HEAD FOR 21000662; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM MODULAR HEAD FOR 21000662; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 75023711
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a thr surgery, two (2) polarstem modular head for 21000662 broke.The procedure was resumed, without any delay, using an equivalent s+n back-up.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
The complaint device used in treatment was not returned for investigation.A product evaluation was not possible.No batch number was communicated so the production history review was not possible.Due to an unknown batch number, the review of historical complaints was performed on product number basis only, revealing 17 additional complaints over the past 12 months with similar failure mode.Due to insufficient information it is not possible to perform a review of past corrective actions.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The performed investigation does not lead to an accurately determined cause.Based on the available information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from repeated resterilization, prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.Therefore, all reusable instruments must be routinely inspected for signs of wear and damage after reprocessing.Any instruments found to be damaged should be replaced as necessary to ensure their proper functioning and to maintain patient safety.This guidance is provided by smith & nephew orthopaedics ag (reference: (b)(4)).There is no need for further actions because of the limited information provided.Nevertheless, smith+nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSTEM MODULAR HEAD FOR 21000662
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24103332
Report Number9613369-2026-00004
Device Sequence Number10312610
Product Code LXH
UDI-Device Identifier07611996114393
UDI-Public7611996114393
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75023711
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Supplement Date Received by Manufacturer02/09/2026
Initial Report FDA Received Date01/16/2026
Supplement Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-