The device was returned to olympus for inspection.Based on the results of the investigation and historical data, reasonably known information suggests probable causes such as electrical problems such as open circuit, short circuit, or signal loss, without any design or manufacturing issue.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
|
The customer reported that, during reprocessing of their bronchovideoscope device, it was discovered that the image produced by the device would black out intermittently, during angulation adjustment.There was no patient involvement.
|