| Model Number |
0285 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This product was manufactured before the implementation of the udi regulation, so the complete udi is unavailable.Applicable us product data has been included in this report.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited high impedance and high pacing threshold outputs due to lead fracture or breakage.This rv lead was explanted, replaced with a different model and returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product was manufactured before the implementation of the udi regulation, so the complete udi is unavailable.Applicable us product data has been included in this report.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited high impedance and high pacing threshold outputs due to lead fracture or breakage.This rv lead was explanted, replaced with a different model and returned for analysis.No additional adverse patient effects were reported.Additional information indicates that the lead fracture was confirmed through fluoroscopy and device-based testing.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of the available information or data determined that the observed gradual rise in low-voltage shock impedance (lvsi) measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in july 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high-voltage shock impedance (hvsi) measurements and reduced shock efficacy.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of fracture was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause traced to device design based on the reported allegation of a gradual rise of shock impedance in a reliance lead system.This performance issue is consistent with capa-8367, which determined this to be a design issue.This product was manufactured before the implementation of the udi regulation, so the complete udi is unavailable.Applicable us product data has been included in this report.
|
| |
|
Event or Problem Description
|
|
It was reported that this right ventricular (rv) lead exhibited high impedance and high pacing threshold outputs due to lead fracture or breakage.This rv lead was explanted, replaced with a different model and returned for analysis.No additional adverse patient effects were reported.Additional information indicates that the lead fracture was confirmed through fluoroscopy and device-based testing.
|
| |
|