Catalog Number 8065752438
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that the leakage occurred from the bottom of cassette before surgery.The procedure type and other surgery details were unknown.It was unknown whether the surgery was completed or not.There was no information about patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The returned wet product was visually inspected and there were no obvious defects observed that would have contributed to the reported event.A pressure leak test was then conducted on the cassette utilizing an external pressure source and no cassette leakage was detected.A calibrated console representing the current software version was then used to functionally test the product.The product could prime and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.No leakage was detected from the manifolds, connectors, drain bag, or drain path between the consumable and console.After functional testing no leakage was detected from the pump elastomer or on the pump area of the fluidics module.We were unable to replicate the customer's experience during laboratory evaluation.The investigation conducted a non-conformance review of the reported lot number.No deviation was identified that would have contributed to this event and all corresponding production release specifications defined in the device master record were met.The presence of fluid leaking from the cassette was not confirmed.After both leakage and console laboratory testing, inspection of the product indicated no signs of continued leakage from the pump elastomer or cassette.Furthermore, there were no signs of fluid leakage onto the fluidics console due to leakage from the cassette.The cassette passed functional and performance testing.The root cause of the customer's complaint could not be established as no fluid leakage was detected during evaluation.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint, therefore no action will be taken for this occurrence.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24105304
Report Number1644019-2026-00389
Device Sequence Number10328892
Product Code LRO
UDI-Device Identifier00380657524389
UDI-Public(01)00380657524389(17)270731(11)250817(10)17LEK9
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752438
Device Lot Number17LEK9
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/25/2025
Supplement Date Received by Manufacturer04/22/2026
Initial Report FDA Received Date01/17/2026
Supplement Report FDA Received Date05/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM.
Patient SexUnknown
-
-