| Catalog Number |
8065753075 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that, during chop mode of the cataract surgery with intraocular lens (iol) implantation, an ophthalmic operating system exhibited low vacuum.The surgery was completed by repriming the cassette without any patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in sections d9, h3, h.6 and h.11.The company representative was able to replicate the reported issue.The nonconforming fluidics module was replaced to address the issue.The system was tested and found to meet product specifications.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event is attributed to the nonconforming fluidics module.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|