Model Number 6688600
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According information received during communication, the internal leakage test failed during pre-use check and replacement of the nozzle unit solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.The nozzle unit is part of the gas module and regulates the inspiratory gas flow to the patient.It consists of a membrane, a mouthpiece and a feather spring.The membrane in the nozzle unit is pressed against a mouthpiece with a feather spring to regulate the gas flow through the gas module.According to the manufacturer¿s specification the complete plastic nozzle unit must be replaced during preventive maintenance.Based on information provided by technician last preventive maintenance was performed on december 2025, which is sooner then a year, or every 5000 hours of operation.The device's logs were not provided for further analysis.The part has not bee returned for further investigation, despite request.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to defective nozzle unit.
 
Event or Problem Description
It was reported that the ventilator failed pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE UNIT SERVO-N
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key24108311
Report Number8010042-2026-0000047
Device Sequence Number18586019
Product Code CBK
UDI-Device Identifier07325710001042
UDI-Public(01)07325710001042(11)191126
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K201874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6688600
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Initial Report FDA Received Date01/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-