• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTIUM USA SUPERLINE II; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTIUM USA SUPERLINE II; DENTAL IMPLANT Back to Search Results
Model Number FXS4512C
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Osseointegration (2646)
Date of Event 12/16/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reported the implant was removed due to osseointegration failure within 1 year, approximately 3 month after implant placement.The x-ray was provided.The bone quality was type iii.The oral hygiene around implant site was moderate.Bone resorption was observed during the removal surgery.Bone augmentation material was used in implant placement surgery.The patient will need another surgical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERLINE II
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DENTIUM USA
6761 katella ave
cypress CA 90630
Manufacturer (Section G)
DENTIUM USA
6761 katella ave
cypress CA 90630
Manufacturer Contact
eun kyung son
6761 katella ave
cypress, CA 90630
7142260229
MDR Report Key24110241
Report Number3005503242-2026-00002
Device Sequence Number10344167
Product Code DZE
UDI-Device IdentifierD755FXS4512C0
UDI-PublicD755FXS4512C0
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160965
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 01/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFXS4512C
Device Catalogue NumberFXS4512C
Device Lot NumberL16MB872S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Initial Report FDA Received Date01/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexMale
-
-