| Model Number |
RT225 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in china that an mr290v autofeed humidification chamber, provided as a part of an rt225 infant ventilator circuit single heated with mr290 autofeed chamber, was found cracked.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) fisher & paykel (f&p) healthcare is currently in the process of finalizing the investigation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the subject mr290v autofeed humidification chamber, provided as a part of an rt225 infant breathing circuit was unable to be returned to f&p healthcare for evaluation as it had been destroyed by the healthcare facility.Our investigation is therefore based on the information and photographs provided, and our knowledge of the product.Results: visual inspection of the provided photographs confirmed the presence of a crack on the chamber dome.Conclusion: based on the information provided by the healthcare facility and without the return of the subject device, we are unable to determine the cause of the reported issue.As part of design validation and verification activities, all f&p healthcare products are tested to ensure they meet documented product requirements and specifications.These requirements and specifications encompass user needs, performance requirements, international product standards as well as quality system requirements.The mr290v vented autofeed humidification chambers are designed and tested to conform to iso 5367 breathing tubes intended for use with anaesthetic apparatus and ventilators.Every mr290v vented autofeed humidification chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v vented autofeed humidification chamber would have met the required specification at the time of production.The user instructions that accompany the rt225 infant breathing circuits state the following: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Do not use the chamber if the seals are not intact when received, or if it has been dropped.Ensure there is a water supply connected to the chamber and that water is present within the chamber.Appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.Visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged.Do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in china that an mr290v autofeed humidification chamber, provided as a part of an rt225 infant ventilator circuit single heated with mr290 autofeed chamber, was found cracked before patient use.There was no patient harm reported.
|
| |
|