|
A distributor reported on behalf of a healthcare facility in china that two mr290v autofeed humidification chambers, provided as a part of an rt225 infant ventilator circuit single heated with mr290 autofeed chamber, were found cracked.There was no patient harm reported.
|
|
(b)(4).Method: the subject mr290v vented autofeed humidification chambers provided with the rt225 infant breathing circuit kit were unable to be returned to f&p healthcare for evaluation as they had been discarded by the healthcare facility.Our investigation is therefore based on the information and photographs provided, and our knowledge of the product.Results: visual inspection of the provided photographs confirmed the presence of a crack on the dome of a chamber.Conclusion: based on the information provided by the healthcare facility and without the return of the subject devices, we are unable to determine the cause of the reported issue.As part of design validation and verification activities, all f&p healthcare products are tested to ensure they meet documented product requirements and specifications.These requirements and specifications encompass user needs, performance requirements, international product standards as well as quality system requirements.All mr290v vented autofeed humidification chambers are visually inspected during the manufacturing process.Additionally, every mr290v vented autofeed humidification chamber is leak tested at the completion of the assembly process.Any chamber which fails these tests is rejected.The subject mr290v vented autofeed humidification chambers would have met the required specification at the time of production.The user instructions that accompany the rt225 infant breathing circuit state the following: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Do not use the chamber if the seals are not intact when received, or if it has been dropped.Ensure there is a water supply connected to the chamber and that water is present within the chamber.Appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.Visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged.Do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.Use usp sterile water for inhalation or equivalent for humidification.Do not add other substances to the water.
|