| Model Number |
RT202 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The subject mr290v vented autofeed humidification chamber provided with an rt202 adult breathing circuit has been requested to be returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in canada reported via a fisher & paykel (f&p) healthcare field representative that an mr290v vented autofeed humidification chamber provided with an rt202 adult breathing circuit was found leaking water after 6 days of patient use.There was no reported patient harm.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in canada reported via a fisher & paykel (f&p) healthcare field representative that an mr290v vented autofeed humidification chamber provided with an rt202 adult breathing circuit was found leaking water after 6 days of patient use.There was no reported patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).Method: the subject mr290v vented autofeed humidification chamber provided with the rt202 adult breathing circuit was returned to fisher & paykel (f&p) healthcare in new zealand, where it was visually inspected and analysed.Results: visual inspection of the returned mr290v vented autofeed humidification chamber confirmed the presence of a crack along the base of the chamber dome.Leak testing confirmed the presence of a leak.Further analysis identified evidence that the chamber had been exposed to significant amounts of an incompatible chemical.Conclusion: the reported water leakage was confirmed and was caused by a crack on the chamber's dome.Based on our investigation, the crack was likely to be caused by exposure to an incompatible chemical, resulting in environmental stress cracking of the chamber dome.Every mr290v vented autofeed humidification chamber is pressure tested following the manufacturing process and any holes, cracks, or leaks would be identified during this process.In addition, the pressure test is followed by a visual inspection of each chamber.Any chamber which fails these tests is rejected.The subject mr290v vented autofeed humidification chamber would have met the required specifications at the time of production.The user instructions for the rt202 adult breathing circuit state the following: - do not soak, wash, or sterilize this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.- use usp sterile water for inhalation or equivalent for humidification.Do not add other substances to the water.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.- ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.- visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damage - do not use the chamber if the seals are not intact when received, or if it has been dropped.
|
| |
|