| Model Number |
407138 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Upon receipt, the device interrogation revealed the battery status eri.The amount of charge taken from the battery was verified.The battery condition was found to be as expected.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.There was no indication of a device malfunction.In conclusion, the pacemaker was fully functional.The battery status was anticipated.
|
| |
|
Event or Problem Description
|
|
Device explanted due to eri indication with unexpected battery behavior.No adverse patient events were reported.Should additional information be received, this file will be updated.
|
| |
|