SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
72201491 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H11 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the device in a plastic bag with one anchor still on the needle, no other anchor can be seen.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.During the investigation it was not possible to identify a definitive root cause; however, it was concluded that the potential causes for the reported event include (1) the application of unintended, inappropriate, or excessive force during device use, and (2) user failure to fully actuate the device during the implantation process.The risk management review confirmed that the reported failure is appropriately documented.The overall risk level is considered adequate.Therefore, no additional actions are deemed necessary at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).H3,h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a meniscus repair surgery, two (2) ultra fast-fix t2 implants failed to deploy.The t1 implant was removed from inside the patient using a grasper.The procedure was resumed using an equivalent s+n backup, after a non-significant delay.No further complications were reported.
|
| |
|