| Model Number |
MR8-T14MH25T |
| Medical Device Problem Codes |
Fracture (1260); Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: no conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in future, product analysis may be performed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the two drill bit was broken and dull and one drill bit was dull.At the time of the report, there was no patient impact.The procedure was completed with backup products and there was no delay in the procedure.It was also reported that there was intervention performed, no damaged piece remained in the patient's body.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the two drill bit was broken and dull and one drill bit was dull.At the time of the report, there was no patient impact.The procedure was completed with backup products and there was no delay in the procedure.It was also reported that there was intervention performed, no damaged piece remained in the patient's body.
|
| |
|
Additional Manufacturer Narrative
|
|
H3:product analysis :evaluation could not be determined that drill bit was broken.However the flutes were observed to be worn.The likely cause of failure is due to flute damage is consistent with a tool contacting another surgical instrument during use leading to reduced cutting performance.It was also noted that cutting efficiency of returned tool was perceived to be slightly less than a comparison tool.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|