| Catalog Number |
25447 |
| Medical Device Problem Codes |
Fracture (1260); Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Therefore, because a serious injury resulted, this event is reportable per 21 cfr part 803.Section h6 was done based on the information provided by the initial reporter and our experience in the investigation of similar complaints.Product return is requested and product will be evaluated after receipt.In case any new or additional information will be gained from this investigation a follow-up report will be sent.Trend is tracked and monitored.
|
| |
|
Event or Problem Description
|
|
Customer reported the fracture of a dental implant and implant loss.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional fda coding being added after investigation of device.Adding additional type of investigation code 10.This is a follow up report to add this additional code.Additional fda coding being added after investigation of device.Adding additional health effect - impact code.4627.This is a follow up report to add this additional code.This is to correct and remove the codes that were initially reported - removing codes for: medical device problem code - 1260.Investigation findings code - 3252 and 3243.The physical investigation of the returned items revealed that no fracture had occurred.The correct codes for this complaint are: medical device problem code - 2408.Investigation findings code - 213.This is a follow up report to correct these/this code(s).
|
| |
|
Search Alerts/Recalls
|