• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS SYS W/CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS SYS W/CONTROL RESERVOIR; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX824T
Medical Device Problem Codes Obstruction of Flow (2423); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 12/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection during the investigation, no significant deformations or damages of the valves were determined.Permeability test a permeability test has shown that both valves are permeable.Computer controlled test to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The result shows that the m.Blue operates not within the accepted tolerances in the vertical position.However, the progav 2.0 operates within the specified tolerances in the horizontal position.An accelerated outflow of m.Blue in the vertical position could be determined.Adjustment test the valves were tested and both valves are adjustable in all pressure settings.Braking force and brake function test the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerance.Internal inspection after dismantling of the valves, organic deposits were found in both valves.Results based on our investigation results, we can determine that there is accelerated outflow at m.Blue.The deposits visible in the valve led to a change in the flow rate.Furthermore, we can determine visible deposits in progav 2.0.The deposits did not affect the technical properties at the time of the investigation.Deposits caused by natural substances found in the body, such as protein, blood, or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic deposits can compromise the integrity of the valve.The examination of the returned product is complete with the creation of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a m.Blue plus (#fx824t) was implanted during a procedure performed on (b)(6) 2025.According to the complainant, the shunt caused underdrainage and had adjustment difficulties.The patient underwent a revision procedure on (b)(6) 2025.The complained product was sent to the manufacturer for investigation.No patient complications were reported as a result of the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS SYS W/CONTROL RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key24114817
Report Number3004721439-2026-00010
Device Sequence Number9515587
Product Code JXG
UDI-Device Identifier04041906504830
UDI-Public(01)04041906504830(11)250717(17)300715(10)20084136
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX824T
Device Catalogue NumberFX824T
Device Lot Number20084136
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2026
Initial Report FDA Received Date01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age10 MO
Patient SexMale
Patient Weight10 KG
-
-