| Catalog Number |
R5C4479C |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a homechoice cassette leaked from an unspecified location.This occurred during fill of peritoneal dialysis therapy.The patient noticed a few drops from the gasket of the cycler.Renal therapy services (rts) advised to end therapy and assisted the patient with removing the supplies.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h6 and h11.H11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|