| Model Number |
PM1240 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Following the electronics performance indicator (epi), an alert was received by the clinician and device was explanted.Patient was stable before, during, and after the procedure and was able to be discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event of electronics performance indicator was not confirmed.The device was above elective replacement indicator (eri) level upon.Analysis of device image and session records indicated no abnormal vorlage drop.Device operated with autocapture and rate responsive features were enabled during implant period.When automatic settings are programmed, such as autocapture and rate responsive feature, the device would adjust itself to accommodate the patient¿s needs for pacing and thus affects the estimated longevity of the device.Further analysis performed did not find any anomaly with battery voltage above elective replacement indicator (eri) level.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
|
| |
|