| Medical Device Problem Codes |
Restricted Flow rate (1248); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron plus g136 they allege that they have low water flow and the handpiece is heating up; no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
01/21/26 evaluation tech: (b)(4).Woe hose, tubing, clogged.No water flow due to a clogged water solenoid, clogged water filter restricting water flow.Worn battery door, blockage in the handpiece cable causing restricted water flow.Will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.No handpiece received for evaluation.Hpc-08220, st-n/a.Dhr review is not required because the product was returned for evaluation, and the described failure mode is a known hazard.
|
| |
|