| Model Number |
379969 |
| Medical Device Problem Codes |
Ambient Noise Problem (2877); Impedance Problem (2950); Unstable Capture Threshold (3269)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
This lead shows noise plus impedance and threshold issues.Lead remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.03-mar-2026 lead was explanted on (b)(6) 2026.Upon receipt, the lead under complaint was found cut 11.5 cm distal to the df4 connector pin (into 2 segments).An extraction aid was found on the distal fragment; it is reasonable to assume that the lead was cut during the surgery.The insulation was found rubbed through 7.5 cm proximal to the lead tip, which is assumed to be the root cause of the reported oversensing in the ventricular channel.Based on the characteristics of this damage, it is reasonable to assume that the lead had been subject to severe mechanical stress in the implanted state.Significant lead motion in the area of the tricuspid valve and interaction with atypical tissues should be taken into consideration.Furthermore, the fixation helix was found bent, which most likely resulted from the extraction procedure due to tensile stress.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|
| |
|