| Catalog Number |
SRLP-490 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".This record is for the left side.The device remains implanted.Unknown if the device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Clarification to b3 date of event: partial date: (b)(6) 2024.Additional, changed, and/or corrected data: b5, b6, d4, g2, h4, h11.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".Healthcare professional confirmed rupture.This record is for the left side.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".Healthcare professional later reported "rupture".This record is for the left side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h.6.
|
| |
|
Event or Problem Description
|
|
Patient reported "rupture".Healthcare professional later reported "rupture".This record is for the left side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6 device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as fold flaw opening and observed a missing piece of shell through microscopic inspection assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases and non penetrating nicks are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Search Alerts/Recalls
|